2020 官网升级中!现在您访问官网的浏览器设备分辨率宽度低于1280px请使用高分辨率宽度访问。
Parameter name: URS-100A, URS-100B, URS-100M
Measuring range: 500~30000mm
Power supply voltage: 24VDC
Output signal: analog quantity: 4~20mADC two-wire system; digital quantity: according to HART communication protocol
Accuracy level: 0.5 level (URS-100A), 0.2 level (URS-100B and URS-100M), intelligent type: 0.2 level
Load capacity: 650Ω (24VDC)
Power supply voltage influence: within the specified voltage range, the output conversion value is less than 0.01%/1V
Long-term stability: ≤±0.3% (URS-100A), ±0.1% (URS-100B and URS-100M)
Compensation temperature range: 0~80°C (URS-100A), -20~80°C (URS-l00B), -20~270°C (URS-100M)
Ambient temperature: -30~80℃
Temperature characteristics: within the temperature compensation range, the error is ≤±0.5% (URS-100A), within the temperature compensation range, ≤±0.1%/10°C (URS-100B and URS-100M)
Response time: ≤1ms
Protection class: IP65
Explosion-proof mark: Ex d II CT4~T6 , Gb Ex ia lI CT1~T6 Ga
Connection form: Flange installation: HG/T20592, PN10 DN40/DN50 RF, or user specified; Bracket installation: 1151 standard pipe bending bracket Screw system installation: G3/4”
Electrical interface: 2 1/2NPT internal threads
Wetted material: cable type: stainless steel, PVC or oil-resistant rubber, fluororubber; rod type: stainless steel, fluororubber
Junction box material: aluminum alloy